Operating Procedures
Sign-in
| 1. Swipe your card through the card reader with the magnetic strip facing the left. |
|
|
| 2. Select which account and instrument to use in order to activate the microscope. |
|
|
 |
 |
 |
 |
 |
 |
 |
Preliminary
| 1. Check that the lamp behind the filament EMISSION-CURRENT is on. If not, contact lab personnel immediately. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2. Ensure that the EDS detector is in the "out" position. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 3. Locate the red power switch for the dSpec control box near the triangular control panel on the left side of the SEM operation console and make sure that the switch is on. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4. Log-in to the computer and open the dPict32 software. Make sure that the stage position reading is X = 25.000 and Y= 35.000. |
|
|
| 5. Check that the stage is set to X = 25 and Y = 35, Tilt = 0, and WD = 39. If these are set wrong, go back to dSpec and select Stage > Home Position. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 6. Turn on the infrared chamber camera and monitor. |
|
|
 |
 |
 |
 |
 |
 |
 |
Specimen Exchange
The procedure is best understood by watching the SEM Specimen Loading Technique video in JEOL JSM-6400 SEM.
Control Panel Settings/ Filament Saturation
The procedure is best understood by watching the SEM Start Up Technique video in JEOL JSM-6400 SEM.
| 1. Turn up the CRT Image contrast knobs and CRT Image brightness knobs on both CRTs. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 2. Depress the MEMO key on the keyboard. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 3. Use the Up or Down arrows on the keyboard to select your choice of accelerating voltage. Press the "L" key for Load. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 4. Make sure you are in the lowest magnification. This can be read in the bottom of the left CRT. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 5. Use the Focus knob to make sure the working distance is 39 mm. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 6. On the control panel, press the SCAN MODE button until the linescan profile (LSP) appears. Depress the SCAN SPEED FAST button to select a fast scan speed. Press the DISPLAY D-MAG button. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 7. Press the ACCEL VOLTAGE. The button should light up. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 8. Saturate the filament slowly by turning the EMISSION knob clockwise. |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
| 9. Turn off DISPLAY D-MAG. To view the image, depress the SCAN PIC button. |
|
|
 |
 |
 |
 |
 |
 |
 |