| 1. Swipe in to the computer logbook to activate the microscope. |
|
|
| 2. Fill cold trap dewar on SEM with 5-6 funnels of liquid nitrogen. |
VIEW |
Click VIEW again to hide.


Click image to open a large version in a new window.
|
| 3. Check that the EDS detector is in the "out" position at 6. |
VIEW |
Click VIEW again to hide.


Click image to open a large version in a new window.
|
| 4. The computer will already be logged on. Do not change the log-in; you cannot log in to your home directory. |
|
|
| 5. Click on the S-4700 icon to start the software. You do not need a password; just click "OK". |
|
|
| 6. The software will display a window with stage coordinates. The proper coordinates are: X=12.5 mm, Y=12.5 mm, and Z=12 mm. Check that the software stage coordinates agree with actual stage position. If not, adjust the stage by hand and make any required changes to the display window. Click "OK". |
VIEW |
Click VIEW again to hide.

Click image to open a large version in a new window.
|
 |
 |
 |
 |
 |
 |
 |